Home » Shop » hot sale » Absolute Process Instruments T16L12S _ transmitter. Fully isolated.

Absolute Process Instruments T16L12S _ transmitter. Fully isolated.

Original price was: $319.20.Current price is: $172.70.

Sale!

72 in stock

SKU: T16L12S Category:

Absolute Process Instruments T16L12S _ transmitter. Fully isolated.

Cecomp Electronics, T16L12S, ThermoPro® Loop Powered Temperature Transmitter, 12″ Length Spring-Loaded for Thermowell Probe

Features:

  • 100 Ω Platinum RTD
  • 316 Stainless Steel Probe
  • 2 wire 4-20 mA Output with Programmable Range
  • -58.0°F to 392.0°F or -50.0°C to 200.0°C
  • NEMA 4X Housing
  • Spring-Loaded Versions for Thermowells
  • User-programmable 4-20 mA output range
  • Selectable °F or °C
  • Pass code protected setup and calibration

Specifications Table:

Range “58.0 to 392.0°F or “50.0 to 200.0°C, selectable
Resolution 0.1°F or 0.1°C resolution
Accuracy Includes linearity error and ±1 LSD 11-point linearization

  • ±0.7°C at “50°C
  • ±0.4°C at 0°C
  • ±0.9°C at 100°C
  • ±1.4°C at 200°C
Display
  • 4 readings per second nominal display update rate
  • 4 digit LCD, 0.5″ H
  • 5 character 0.25″ H alphanumeric
  • Low battery voltage indication(ALARM1 under range indication on display, ALARM2 over range indication on display)
Sensor
  • IEC-751 Class B 100 Ω Platinum RTD, 0.00385 alpha curve 1 /2″ NPT male, 316L stainless steel
  • All wetted parts are 316L stainless steel
  • Spring-loaded probe (Spring-loaded probe(thermowell required for spring-loaded versions, spring-loaded probe fits standard thermowells)
Analog Output
  • 12,000 count 4-20 mA output
  • Updated approximately 4 times per second
  • Temperature range configurable to correspond to 4-20 mA output
  • Configurable for upscale or downscale burnout
Controls and Functions
  • 3 button front keypad for setup, calibration, and configuration.
  • User-defined pass codes to prevent unauthorized changes.
  • Pass code required for configuration and calibration.
Output Test Function When TEST button is held loop current and display are set to test level, independent of temperature input, to allow testing of system operation. Test level is user set. Output test function can be pass code protected
Calibration
  • User settable pass code required to enter calibration mode
  • Zero and span temperature calibration
  • Non-interactive zero, span, and linearity, ±10% of range
Weight
  • Product: 12 ounces (approximately)
  • Shipping: 1 pound (approximately)
Housing Materials ABS/polycarbonate NEMA 4X case, rear gasket, polycarbonate label, aluminum sensor block
Storage Temperature “40 to 203°F (“40 to 95°C)
Operating Range “4 to 185°F (“20 to 85°C) at housing
Cable Length 12″
Power
  • Any DC supply/loop resistance that maintains 8 to 32 VDC at gauge terminals
  • Reverse polarity protected
  • 3 ft long, 2-conductor 22 AWG cable